DDS Sine wave generator/output
Frequency range of sine wave generator = 10Hz to 4MHz
DDS sine wave generator
Sine wave quality = 62dB SNAD (nominal)
Output voltage = 2Vpp maximum
Output attenuator, 0dB to -20.0dB in 2.5dB steps
DC offset at output = 0.9mV typical
Output impedance = 50 Ohms + parasitics of 23nH and 33pF (Z mode)
Frequency accuracy at 1MHz is ±100ppm
BNC output connector
***** Destructive/Absolute Maximum Limits *****
Absolute maximum current applied to output = 0.2Arms
Absolute maximum voltage applied to output = 9VDC or 22VppAC

Input channel
Input impedance = 1M ohm in parallel with 14.5pF
AC coupled input (0.15Hz HP)
Maximum signal input = 3.5Vpp maximum before compression (+5dB)
Minimum input signal = 100uVpp (-86dB) with Rsource = 50R
Noise floor = -86dB at 1MHz/input S/C; -73dB at 1MHz/input O/C
BNC input connector
Absolute maximum voltage applied to input = 50VDC or 50VppAC

Other hardware features
USB connection for data and power USB 2.0 full speed (12Mbps)
Mode led indicating Power and USB activity
Current consumption = 100mA to 220mA depending on activity
Power consumption = 0.5Watt to 1.1Watt depending on activity

Other details
Advanced graphing program used to display test results
Maximum number of test points = 1000 per plot, 2 parameters per plot
Multiple plots; 10 plots per graph, with user text fields
Logarithmic and linear frequency sweep and stationary frequency
Logarithmic and linear frequency axis
User controllable test settling time
User selectable range of frequency points
Logarithmic (dB) Gain/Loss amplitude display
Amplitude display range = +5dB to –86dB (0.025dB best resolution)
Amplitude detector straight line deviation = ±0.35dB
Phase accuracy = see graph in User Manual
Phase range = ±180o and 0o to -180o
|Z| The modulus of the impedance of the test device (DUT)
Impedance accuracy = from ±0.5% to ±4.5% deviation (see graphs)
Impedance displayed in Ohms or Admittance in Siemens (log & linear)
Impedance display range of 0.01 Ohms to 1M Ohm (600K Ohm)
Auto ranging or fixed Impedance and Gain axes
Save/Recall as GAD and GZD files
Save graphs as JPEG images, BMP images, Meta files and data files
Use graphs in documents and emails
Graphing Controls; START, STOP, REPEAT, ZOOM, UNDO ZOOM, FULL SCREEN, USER TEXT FIELDS, FREQUENCY LIMITS, SWEEP BACKWARDS, ALTERNATE SWEEP, SWEEP TO ZOOM EXTENTS, PEN COLOURS, PEN LOCK, GRAPH SELECTION COLOUR/BLACK/BLACK PLUS MARKERS, COPY, CUT, PASTE
Calibration option for internal amplitude and phase trims
Multiple document operation
Scripting control of the unit

PC requirements
Operating systems: Microsoft Windows 98SE, ME, 2000 and XP
Pentium 500MHz, 64Mbytes of RAM, 50Mbytes of free hard disk space, mouse, keyboard, USB port, graphics card with 1024x768 resolution.

(return to main products page)